CELL-DYN Ruby

Granulocytosis

Description

Both cases (A and B) show a significant granulocytosis (neutrophilia). For Case A, the neutrophil cluster retains a relatively normal shape but its position is slightly higher on the SIZE axis compared to normal. The sample also shows monocytosis, moderate normocytic anaemia and a high normal platelet count. The BAND flag suggests the possibility of a left shift.

 

For comparison, Case B (51 year old male following nephrectomy for renal cell carcinoma) shows a less marked neutrophilia, with relatively normal RBC and platelet results. With regards to the graphical displays, the neutrophil population cluster (upper left plot) is also relatively normal but the extended location of the cluster along the LOBULARITY axis in the other WBC plot (upper right) indicates that the neutrophils are hypersegmented.

Numerical Results

 
CASE A CASE B

WBC  36.5 10e3/μL  16.3 10e3/μL
NEU 34.6 94.9 %  BAND 13.4 82.0 %
MONe 1.39 3.82 % .835 5.12 %
EOS .011 .030 % .025 .151 %
BASO .047 .130 % .057 .351 %
LYMe .409 1.12 % 2.02 12.4 %

 
RBC  2.70 10e6/μL  4.22 10e6/μL
HGB 7.79 g/dL  13.1 g/dL
HCT  23.9 % 39.6 %
MCV 88.5 fL 93.8 fL
MCH 28.8 pg 31.1 pg
MCHC 32.6 g/dL 33.1 g/dL
RDW 13.4 % 11.0 %

 
PLT 461. 10e3/μL 229. 10e3/μL
MPV 6.44 fL  9.78 fL

 
Granulocytosis:
CASE A
Granulocytosis:
CASE B
CELL-DYN Ruby Casebook. Granulocytosis. 2018. ADD-00005070. p.35
©2021 Abbott, Abbott Park, Illinois, U.S.A.

Privacy Policy| Terms and Conditions

Unless otherwise specified, all product and service names appearing in this Internet site are trademarks owned by or licensed to Abbott, its subsidiaries or affiliates. No use of any Abbott trademark, trade name, or trade dress in this site may be made without the prior written authorization of Abbott, except to identify the product or services of the company.

All Alinity hq, CELL-DYN Ruby and CELL-DYN Sapphire instruments are Class I laser products. ACCELERATOR a3600 is a Class II laser product. 

The Alinity, ACCELERATOR and CELL-DYN systems are intended for performing in vitro diagnostic assays on samples of human origin (blood, urine). Read the instructions in the system manuals and labeling and/or reagent instructions carefully. Manufacturer or Authorized Representative: Abbott Germany.

Alinity h-series is available in select countries, not including the US.