CELL-DYN Ruby

Lymphocytosis

Description

This patient had lymphocytosis, accompanied by a VAR LYM flag. In the WBC Differential plot the lymphocytes and monocytes seemed to form a single coherent population. The SIZE/ COMPLEXITY and the 90º/WBC 0º scatterplot demonstrated that part of the lymphocytes were classified as monocytes. It were actually the larger lymphocytes that were considered as monocytes, fully in agreement with the morphological observation of atypical, monocytoid lymphocytes.

 

This combination of findings is highly characteristic of Infectious Mononucleosis, a diagnosis that was confirmed in this patient with serological testing.

Numerical Results

 
SUSPECT
WBC 22.0 10e3/μL
NEU 2.633 12.006 %
BLST .056 .255 %
MONe .203 .923 %
EOS .084 .382 %
BASO .091 .414 %
LYM 18.97 86.09 %  VAR LYM

 
RBC 4.70 10e6/μL
HGB 12.4 g/dL
HCT 37.4 %
MCV 79.5 fL
MCH 26.3 pg
MCHC 33.1 g/dL
RDW 12.9 %

 
PLT 374. 10e3/μL
MPV  4.22 fL

 
:
:
:
:
:
:
CELL-DYN Ruby Casebook. Lymphocytosis. 2018. ADD-00005070. p.39
©2021 Abbott, Abbott Park, Illinois, U.S.A.

Privacy Policy| Terms and Conditions

Unless otherwise specified, all product and service names appearing in this Internet site are trademarks owned by or licensed to Abbott, its subsidiaries or affiliates. No use of any Abbott trademark, trade name, or trade dress in this site may be made without the prior written authorization of Abbott, except to identify the product or services of the company.

All Alinity hq, CELL-DYN Ruby and CELL-DYN Sapphire instruments are Class I laser products. ACCELERATOR a3600 is a Class II laser product. 

The Alinity, ACCELERATOR and CELL-DYN systems are intended for performing in vitro diagnostic assays on samples of human origin (blood, urine). Read the instructions in the system manuals and labeling and/or reagent instructions carefully. Manufacturer or Authorized Representative: Abbott Germany.

Alinity h-series is available in select countries, not including the US.