CELL-DYN Ruby

Platelet Aggregation

Description

This sample was a more distinct example of platelet aggregation. The NWBC flag was set and was accompanied by an URI flag, indicating interference in the region between the PLT and RBC populations. This interference was clearly visible in the 0º/RBC 10º scatterplot as a cluster of black events above the platelets and was strongly suggestive for the presence of platelet aggregation.

 

When a low platelet count is detected in combination with the NWBC and URI flag, it is advised to check the sample for the presence of platelet aggregation.

Numerical Results

 
  SUSPECT
WBC 2.29 10e3/μL
NEU 1.279 55.816 %
BLST .001 .039 % NBWC
MONe .445 19.5 %
EOS .082 3.57 %
BASO .034 1.48 %
LYM .450 19.634 %

 
RBC 2.63 10e6/μL
HGB 7.88 g/dL
HCT 23.2 %
MCV 88.2 fL  RBC MORPH
MCH 30.0 pg
MCHC 34.0 g/dL
RDW 19.8 %

 
PLT 65.2* 10e3/μL
MPV 10.3* fL  URI

 
:
:
:
:
:
:
CELL-DYN Ruby Casebook. Platelet Aggregation. 2018. ADD-00005070. p.48
©2021 Abbott, Abbott Park, Illinois, U.S.A.

Privacy Policy| Terms and Conditions

Unless otherwise specified, all product and service names appearing in this Internet site are trademarks owned by or licensed to Abbott, its subsidiaries or affiliates. No use of any Abbott trademark, trade name, or trade dress in this site may be made without the prior written authorization of Abbott, except to identify the product or services of the company.

All Alinity hq, CELL-DYN Ruby and CELL-DYN Sapphire instruments are Class I laser products. ACCELERATOR a3600 is a Class II laser product. 

The Alinity, ACCELERATOR and CELL-DYN systems are intended for performing in vitro diagnostic assays on samples of human origin (blood, urine). Read the instructions in the system manuals and labeling and/or reagent instructions carefully. Manufacturer or Authorized Representative: Abbott Germany.

Alinity h-series is available in select countries, not including the US.