CORE LABORATORY
Don't have an account? Sign Up

CELL-DYN Ruby

URI Flag

Description

The URI flag in this specimen indicated that there was an abnormality detected in the upper region of the platelet cluster. The 0º/RBC 10º scatterplot demonstrated the presence of large platelets and this was in agreement with the MPV of 10.0 fL. The RBC are microcytic and hypochromic, but despite the small volume of the RBC, the separation between the RBC and PLT cluster was still perfect.

 

This picture may be compatible with reactive platelets that are generally larger than normal, in combination with a disorder of microcytic RBC, probably thalassemia trait.

Numerical Results

 
WBC 7.20 10e3/μL
NEU 4.179 58.081 %
BLST 0.00 0.00 %
MONe .520 7.22 %
EOS .179 2.48 %
BASO .090 1.26 %
LYM 2.223 30.93 %

 
RBC 5.84 10e6/μL
HGB 10.5 g/dL
HCT 35.6 %
MCV 61.0 fL  RBC MORPH
MCH 18.0 pg
MCHC 29.6 g/dL
RDW 12.8 %

 
PLT 221.* 10e3/μL
MPV 10.0* fL  URI

 
:
:
:
:
:
:
CELL-DYN Ruby Casebook. URI Flag. 2018. ADD-00005070. p.47
©2021 Abbott, Abbott Park, Illinois, U.S.A.

Privacy Policy| Terms and Conditions

Unless otherwise specified, all product and service names appearing in this Internet site are trademarks owned by or licensed to Abbott, its subsidiaries or affiliates. No use of any Abbott trademark, trade name, or trade dress in this site may be made without the prior written authorization of Abbott, except to identify the product or services of the company.

All Alinity hq, CELL-DYN Ruby and CELL-DYN Sapphire instruments are Class I laser products. ACCELERATOR a3600 is a Class II laser product. 

The Alinity, ACCELERATOR and CELL-DYN systems are intended for performing in vitro diagnostic assays on samples of human origin (blood, urine). Read the instructions in the system manuals and labeling and/or reagent instructions carefully. Manufacturer or Authorized Representative: Abbott Germany.

Alinity h-series is available in select countries, not including the US.